COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 2993788 · Received March 8, 2013

Report

Report Number
1061932-2013-00374
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 8, 2013
Report Date
February 8, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED A SMALL AMOUNT OF DRIED BLOOD AT THE BOTTOM OF THE PIERCING MODULE AND CLEANED UP SPILLAGE UTILIZING PERSONAL PROTECTIVE EQUIPMENT (PPE). THE FSE REPLACED THE ASPIRATION TUBING FROM THE NEEDLE TO THE BLOOD DETECTOR (BD1). THE FSE NOTED SLOW FLOW THROUGH THE BLOOD SAMPLING VALVE (BSV) AND DISASSEMBLED AND CLEANED THE BLOOD SAMPLING VALVE. THE FSE PERFORMED SYSTEM STARTUP AND SYSTEM VERIFICATION AND COMPLETED RETAINED PATIENT SAMPLES WITHOUT ERRORS. THE FSE VERIFIED NO FLUID LEAK AT THE NEEDLE AND PIERCING MODULE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY FIFTY MILLILITERS OF FLUID LEAKED OUTSIDE THE INSTRUMENT FROM THE SAMPLE LINE THAT HAD COME OFF THE NEEDLE ASSEMBLY INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT SAMPLES WERE NOT ANALYZED AT THE TIME OF THE EVENT. PATIENT RESULTS WERE NOT IMPACTED. THE LABORATORY HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
100229 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA