COULTER® ACT DIFF 12¿ ANALYZER

MDR report key: 2993787 · Received March 8, 2013

Report

Report Number
1061932-2013-00373
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 8, 2013
Report Date
February 8, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K973634
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) REPLACED FAULTY PINCH VALVE LV13 AND RESOLVED THE ISSUE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY FIVE MILLILITERS OF CLEAR DILUENT LEAKED UNDER THE INSTRUMENT INVOLVING THE COULTER ACT DIFF 12 ANALYZER. THE OPERATOR WAS WEARING PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
99738 COULTER® ACT DIFF 12¿ ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA