FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 2993786 · Received March 8, 2013

Report

Report Number
2955842-2013-00770
Event Type
Malfunction
Date Received
March 8, 2013
Date of Event
February 7, 2013
Report Date
February 7, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN LOCATED AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND IT WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. ADDITIONAL OBSERVATIONS NOT REPORTED BY THE SITE WERE INDENTATIONS AND SCRATCHES MARKS ON EDGES AND SURFACE OF THE DISTAL PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DAVINCI SI SURGICAL PROCEDURE, THE CUSTOMER NOTED A BROKEN WIRE ON THE LARGE NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98744 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-06 M10120416 697

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES