FDA Adverse Event Malfunction Summary report: N

FIX-SLEEVE

MDR report key: 2993699 · Received March 7, 2013

Report

Report Number
1719045-2013-00493
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
February 1, 2013
Report Date
February 11, 2013
Manufacturer
SYNTHES MONUMENT
Product Code
FZX
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.

Additional Manufacturer Narrative · 1

THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. ISIMAC MACHINE COMPANY MANUFACTURED THE DRILL STOP, PART 357.405, AND LOT NUMBER 6623811. THE MATERIAL OF THE HOUSING AND THE BUTTON WERE CONFIRMED TO BE CORRECT. FUNCTIONAL TESTING WAS PERFORMED USING GAGE GT026298 AND THE DRILL STOP PASSED THE FUNCTIONAL TEST. THE MATERIAL WAS TESTED AND CONFORMED TO SPECIFICATIONS. THE INSTRUMENT CONFORMED TO DIMENSIONAL SPECIFICATIONS AT THE TIME OF MANUFACTURING AND PASSED FUNCTIONAL TESTING AT SYNTHES INCOMING INSPECTION.

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. REVIEW OF THE DEVICE HISTORY RECORDS SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT IS REPORTED THAT ON (B)(6) 2013, THE FIX-SLEEVE WOULD MOVE BACKWARDS IF IT WAS MINIMALLY TILTED. IT WOULD HOLD WHEN POSITIONED STRAIGHT, BUT IF IT WAS TILTED AT ALL, IT WOULD MOVE BACKWARDS DURING DRILLING. THIS IS 1 OF 1 REPORTS FOR THIS EVENT.

Description of Event or Problem · 1

THIS IS 1 OF 1 REPORT FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98107 FIX-SLEEVE FZX SYNTHES MONUMENT 6623811

Patients

Seq Age Sex Outcome Treatment
1