PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2993635 · Received March 7, 2013

Report

Report Number
3004209178-2013-91639
Event Type
Injury
Date Received
March 7, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST AND MOTOR ERROR ALARMED DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT HAD MISSING END CAP STICKER, CRACKED DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, BROKEN BELT CLIP SLOT AND SCRATCHED DISPLAY WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS IN THE HOSPITALIZED RECOVERING FROM A C-SECTION, AND THE BLOOD GLUCOSE WAS 16MMOL/L. TROUBLESHOOTING WAS PERFORMED. IT WAS STATED THAT THE INSULIN PUMP ALARMED, AND THE DRIVE SUPPORT CAP WAS PROTRUDED. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98273 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722CAL

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization