PARADIGM INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-91646
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- February 25, 2013
- Report Date
- February 25, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- K040676
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND
- Reporter Occupation
- PATIENT
Narratives
PRIME ALARM DURING BASIC OCCLUSION TEST DUE TO LOOSE/PROTRUDING DRIVE SUPPORT DISK. CRACKED RESERVOIR TUBE LIP AND CRACKED CASE NEAR RESERVOIR WINDOW NOTED DURING VISUAL INSPECTION.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
THE CUSTOMER REPORTED THAT INSULIN SQUIRTED OUT WHILE PRIMING AND THE DEVICE ALARMED. THE BLOOD GLUCOSE READING WAS 165MG/DL. THE CUSTOMER STATED THAT SHE ATTEMPTED TO REWIND SEVERAL TIMES WITHOUT RESULTS. TROUBLESHOOTING WAS PERFORMED. THE CALLER STATED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. ADVISED THE CUSTOMER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98279 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-715NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |