FDA Adverse Event Injury Summary report: N

TECNIS 1 MULTIFOCAL

MDR report key: 2993584 · Received March 7, 2013

Report

Report Number
9614546-2013-00046
Event Type
Injury
Date Received
March 7, 2013
Date of Event
November 29, 2012
Report Date
February 15, 2013
Manufacturer
ABBOTT MEDICAL OPTICS
Product Code
MFK
PMA / PMN Number
P980040
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT MEDICAL OPTICS HAS BEEN SUBMITTED. PLACEHOLDER.

Additional Manufacturer Narrative · 1

THE EXPLANTED INTRAOCULAR LENS (IOL) WAS RETURNED TO OUR MANUFACTURING FACILITY FOR EVALUATION. VISUAL INSPECTION SHOWED A LENS WHERE THE OPTIC WAS CUT IN HALF, WHICH IS CONSISTENT WITH A LENS THAT HAS BEEN REMOVED FROM THE EYE. DUE TO THE RECEIVED CONDITION OF THE LENS, DIOPTER MEASUREMENT COULD NOT BE PERFORMED. NO OPTICAL DEVIATIONS ON THE OPTIC PARTS WERE FOUND. THE LENS PASSED ALL ACCEPTANCE CRITERIA FOR ALL TESTS PERFORMED AND WAS WITHIN ALL SPECIFICATIONS DURING THE MANUFACTURING PROCESS. (B)(4): PLACEHOLDER.

Additional Manufacturer Narrative · 1

MANUFACTURING RECORD REVIEW REVEALED THAT THE DIOPTER MEASUREMENT RECORDS FROM THIS PARTICULAR LENS WERE VERIFIED AND SHOWN TO BE WITHIN SPECIFICATIONS. THIS IS IN COMPLIANCE WITH THE LABELED DIOPTRIC POWER 22.5 DIOPTER OF THIS LOT NUMBER. IN CONCLUSION, THE BATCH HAS PASSED ALL ACCEPTANCE CRITERIA FOR ALL TESTS PERFORMED AND IS WITHIN ALL SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT AN IMPLANTATION OF AN INTRAOCULAR LENS (IOL) WHICH WAS EXPLANTED IN A SECONDARY PROCEDURE DUE TO AN UNEXPECTED MYOPIC OUTCOME. IT WAS STATED THAT THERE WAS NO INCISION ENLARGEMENT MADE DURING REMOVAL AND EXCHANGE OF THE LENS. NO COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97001 TECNIS 1 MULTIFOCAL MULTIFOCAL IOLS MFK ABBOTT MEDICAL OPTICS ZMB00

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention