RELIANCE SYNERGY WASHER

MDR report key: 2993261 · Received March 7, 2013

Report

Report Number
9680353-2013-00023
Date Received
March 7, 2013
Date of Event
February 7, 2013
Report Date
March 7, 2013
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED AT THE FACILITY AND FOUND THE RECIRCULATION PUMP LEAKING. THE CAUSE OF THE LEAK WAS A DETERIORATED PUMP GASKET. THE UNIT WAS INSTALLED IN 2010 AND HAS NOT BEEN UNDER STERIS WARRANTY OR SERVICE CONTRACT SINCE (B)(6) 2011. TO RESOLVE THE ISSUE, THE TECHNICIAN REPLACED THE PUMP GASKET, TESTED THE UNIT AND CONFIRMED THE UNIT WAS OPERATIONAL.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THE RELIANCE WASHER WAS LEAKING WATER ONTO THE FLOOR AND THROUGH MULTIPLE ROOMS. THERE WERE NO PROCEDURAL DELAYS/CANCELLATIONS OR INJURIES REPORTED.