FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2993254 · Received March 7, 2013

Report

Report Number
1416980-2013-05531
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
December 15, 2012
Report Date
February 14, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS A FALSE EMPTY DETECT AND USE ERROR WITH THE INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED. THE HOMECHOICE/HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE STATES IN THE WARNINGS: "SETTING THE I-DRAIN ALARM VOLUME TOO LOW OR OFF CAN RESULT IN AN INCOMPLETE INITIAL DRAIN FOLLOWED BY A FULL FILL. THIS CAN RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION. SEE TABLE 9-1 ON PAGE 9-10 FOR THE RECOMMENDED STARTING POINTS WHEN DETERMINING YOUR OPTIMAL I-DRAIN ALARM VOLUME" AND TABLE 9-1 ON PAGE 9-10 GIVES THE RECOMMENDED STARTING POINT FOR I-DRAIN ALARM SETTING AS 70% OF THE LAST FILL VOLUME. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2012 AT 23:25:49. DURING NIGHT DRAIN CYCLE ONE, THE PATIENT'S ULTRAFILTRATION READING WAS 3056ML, INDICATING THE HOME PATIENT (HP) DRAINED 3056ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2700ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97995 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 57 YR