FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2993044 · Received March 7, 2013

Report

Report Number
2517506-2013-00042
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
January 10, 2013
Report Date
February 14, 2013
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
KXT
PMA / PMN Number
K093441
Removal / Correction Number
2517506-03-06-2013-003-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS CONFIRMED CUSTOMER COMPLAINTS OF LOW QC AND PATIENT RECOVERY FOR DIGOXIN WHEN THIS CALIBRATOR LOT IS USED TO CALIBRATE THE DIGXN METHOD. INTERNAL TESTING HAS CONFIRMED AN AVERAGE 15% LOW BIAS IN THE THERAPEUTIC RANGE OF 0.9 TO 2.0 NG/ML. THE CAUSE OF THE BIAS LOW IN DIGOXIN RECOVERIES IS CALIBRATOR INACCURACY. SIEMENS HEALTHCARE DIAGNOSTICS INC. CONDUCTED A VOLUNTARY RECALL FOR DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053. AN URGENT MEDICAL DEVICE RECALL, COMMUNICATION #(B)(4), WAS ISSUED IN (B)(4) 2013 TO IMPACTED CUSTOMERS. CUSTOMERS WERE INSTRUCTED TO IMMEDIATELY DISCONTINUE USE AND TO DISCARD ANY REMAINING INVENTORY OF DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053.

Description of Event or Problem · 1

BIASED LOW DIGOXIN RESULTS WERE OBTAINED ON CALIBRATION AND QC IMMEDIATELY AFTER CALIBRATION WITH A NEW LOT OF CALIBRATOR. IT IS UNKNOWN IF PATIENT RESULTS WERE REPORTED TO THE PHYSICIAN. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED AS A RESULT OF THE FALSELY DEPRESSED DIGOXIN RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED DIGOXIN RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98062 DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM DRUG 4 CALIBRATOR KXT SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW 2KD052

Patients

Seq Age Sex Outcome Treatment
1