DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00040
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- January 11, 2013
- Report Date
- February 14, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- KXT
- PMA / PMN Number
- K112142
- Removal / Correction Number
- 2517506-03-06-2013-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS CONFIRMED CUSTOMER COMPLAINTS OF LOW QC AND PATIENT RECOVERY FOR DIGOXIN WHEN THIS CALIBRATOR LOT IS USED TO CALIBRATE THE DIGXN METHOD. INTERNAL TESTING HAS CONFIRMED AN AVERAGE 15% LOW BIAS IN THE THERAPEUTIC RANGE OF 0.9 TO 2.0 NG/ML. THE CAUSE OF THE BIAS LOW IN DIGOXIN RECOVERIES IS CALIBRATOR INACCURACY. SIEMENS HEALTHCARE DIAGNOSTICS INC. CONDUCTED A VOLUNTARY RECALL FOR DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053. AN URGENT MEDICAL DEVICE RECALL, COMMUNICATION #(B)(4), WAS ISSUED IN (B)(4) 2013 TO IMPACTED CUSTOMERS. CUSTOMERS WERE INSTRUCTED TO IMMEDIATELY DISCONTINUE USE AND TO DISCARD ANY REMAINING INVENTORY OF DIMENSION VISTA DRUG 4 CAL (KC460) LOT 2KD052 AND DRUG 4 CAL (KC460A) LOT 2KD053.
BIASED LOW DIGOXIN RESULTS WERE OBTAINED ON CALIBRATION AND QC IMMEDIATELY AFTER CALIBRATION WITH A NEW LOT OF CALIBRATOR. THE CALIBRATION WAS NOT ACCEPTED . PATIENT RESULTS WERE NOT REPORTED TO THE PHYSICIAN FOR THIS CALIBRATION. AFTER THE SHIFT LOW WAS NOTED BY THE LABORATORY, THE CUSTOMER WAS SENDING THEIR DIGOXIN TESTING TO AN OUTSIDE LABORATORY. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED AS A RESULT OF THE FALSELY DEPRESSED DIGOXIN RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED DIGOXIN RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98061 | DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM | DRUG 4 CALIBRATOR | KXT | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 2KD053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |