Narratives

Additional Manufacturer Narrative · 1

(B)(4). REFRACTIVE SURPRISE. DEVICE EVALUATED BY MANUFACTURER? NO - LENS NOT RETURNED. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION METHOD: WORK ORDER SEARCH, MEDICAL REVIEW. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND NO VISIBLE DAMAGE TO THE LENS. THE LENS WAS RETURNED IN LIQUID. THE LENS WAS REHYDRATED IN BSS FOR RE-MEASUREMENT. THE LENS LENGTH, WIDTH AND DIOPTER WERE MEASURED AND THE RESULT OF THE MEASUREMENTS WAS COMPARED AGAINST THE ORIGINAL VALUES AND THE LENS WAS FOUND TO BE IN SPECIFICATION. MEDICAL REVIEW - REVIEW OF THE FILE INDICATES THAT THE LENS WAS NOT IMPLANTED IN ACCORDANCE WITH THE DFU REQUIREMENTS ((B)(6), ACD BELOW 2.8MM FOR ICL/TICL AND BELOW 3.0MM FOR VICL/VTICL, KERATOCONUS, PREGNANT OR NURSING PATIENTS, PATIENTS WITH LOW/ABNORMAL CORNEAL ENDOTHELIAL CELL DENSITY, FUCH'S DYSTROPHY OR OTHER CORNEAL PATHOLOGY, PATIENTS WHO ARE AMBLYOPIC OR BLIND IN THE FELLOW EYE). OFF-LABEL USE OF THE DEVICE, NO CLINICAL DATA CAN SUPPORT THE COMPLAINT EVENT(S) OR THE EFFECT(S) ON THE EFFICACY AND SAFETY OF THE DEVICE. THIS INFORMATION HAS BEEN COMMUNICATED TO THE MEDICAL ADVISOR OF STAAR SURGICAL AND TO THE SURGEON IN CASE OF SEVERE DETERIORATION OF PATIENT HEALTH. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH, MEDICAL REVIEW AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.6MM VICM 12.6 IMPLANTABLE COLLAMER LENS IN THE PATIENT'S LEFT EYE (OS) ON (B)(6) 2012. THE PATIENT EXPERIENCED A REFRACTIVE SURPRISE AND THE LENS WILL BE EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97890 VISIAN ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS MTA STAAR SURGICAL COMPANY VICM 12.6 NA

Patients

Seq Age Sex Outcome Treatment
1 38 YR Required Intervention CARTRIDGE MODEL AND LOT NUMBER UNK| INJECTOR MODEL AND LOT NUMBER UNK