HI-TORQUE WHISPER ES GUIDE WIRE
Report
- Report Number
- 2024168-2013-01318
- Event Type
- Injury
- Date Received
- March 7, 2013
- Date of Event
- February 13, 2013
- Report Date
- February 13, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- DQX
- PMA / PMN Number
- K101116
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). MANUFACTURING SITE CORRECTED. EVALUATION SUMMARY: THE COMPLAINT DEVICE WAS RETURNED AND THE REPORTED SEPARATION WAS CONFIRMED VIA RETURNED DEVICE ANALYSIS. BASED ON VISUAL ANALYSIS AND SCANNING ELECTRON MICROSCOPY (SEM) IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. ADDITIONALLY, A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NONCONFORMANCES WITH THE REPORTED LOT AND A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
IT WAS REPORTED THAT THE PATIENT HAD SEVERE LEFT MAIN, OSTIAL LEFT ANTERIOR DESCENDING ARTERY (LAD), OSTIAL CIRCUMFLEX (CX) AND DISTAL CX DISEASE. THE TARGET AREAS WERE REPORTED TO BE HEAVILY CALCIFIED. THE WHISPER GUIDE WIRE TIP SEPARATED DURING USE. THE SEPARATED GUIDE WIRE TIP WAS ATTEMPTED TO BE SNARED. HOWEVER, THE ATTEMPT WAS UNSUCCESSFUL. THE SEPARATED TIP REMAINS IN THE PATIENT. THERE WAS GOOD FLOW THROUGHOUT THE PROCEDURE AND NO PERFORATION OR FLOW DISRUPTION WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96939 | HI-TORQUE WHISPER ES GUIDE WIRE | GUIDE WIRES | DQX | AV-TEMECULA-CT | 2100871 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 91 YR | Other| R |