FDA Adverse Event Injury Summary report: N

SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

MDR report key: 2992676 · Received March 7, 2013

Report

Report Number
1416980-2013-05487
Event Type
Injury
Date Received
March 7, 2013
Date of Event
January 22, 2013
Report Date
February 11, 2013
Manufacturer
BAXTER HEALTHCARE
Product Code
KDJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A SAMPLE WAS NOT REQUESTED AS THIS EVENT INVOLVED USE ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THIS REPORT OF PERITONITIS CAUSED BY A BREACH IN ASEPTIC TECHNIQUE WAS CONFIRMED. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT HAD A BREAK IN ASEPTIC TECHNIQUE DURING PERITONEAL DIALYSIS (PD) THERAPY, WHICH CAUSED PERITONITIS. THE PATIENT HAD A BREAK IN ASEPTIC TECHNIQUE FURTHER DESCRIBED AS THE PATIENT MADE MISTAKE/TOUCH CONTAMINATION AND DID NOT CLEAN THE EXCHANGE AREA BEFORE STARTING PD. THE PATIENT EXPERIENCED PERITONITIS AND WAS HOSPITALIZED FOR THE EVENT. TREATMENT RENDERED WAS NOT REPORTED. PD THERAPY WAS ONGOING. THE PATIENT WAS RECOVERING FROM THE EVENT OF PERITONITIS. ADDITIONAL INFORMATION WAS REQUESTED BUT IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97799 SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 58 YR Hospitalization DIANEAL 2.5% LOW CALCIUM AMBFLEX| DIANEAL 1.5% LOW CALCIUM AMBUFLEX| EXTRANEAL VIAFLEX| HOMECHOICE| DIANEAL 1.5% LOW CALCIUM ULTRABAG