FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 2992598 · Received March 7, 2013

Report

Report Number
3008382007-2013-04350
Event Type
Malfunction
Date Received
March 7, 2013
Report Date
February 7, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE LAY USER/PATIENT IN USA CONTACTED LIFESCAN TO REPORT THE ONE TOUCH PING METER HAD A DISPLAY ISSUE WITH A LINE THROUGH THE DISPLAY. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96878 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3306968

Patients

Seq Age Sex Outcome Treatment
1