FDA Adverse Event
Malfunction
Summary report: N
ORTHO SUMMIT SAMPLE HANDLING SYSTEM
MDR report key: 2992591
·
Received March 7, 2013
Report
- Report Number
- 2250051-2013-00058
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- February 28, 2013
- Report Date
- March 7, 2013
- Manufacturer
- HAMILTON BONADUZ AG
- Product Code
- MMH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE FE INVESTIGATED THE INSTRUMENT FOR THE REPORTED ISSUE. THE FE INSPECTED ALL TIP AND PLUNGER CLAMPS. NO REPAIRS WERE PERFORMED. THE FE PERFORMED SPLLD CHECKING PROCEDURE AND TESTED THE SUMMIT WITH OAS USER SOFTWARE. THE INSTRUMENT WAS TESTED WITHOUT PROBLEM.
Description of Event or Problem · 1
THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT REPORTEDLY DID NOT PIPETTE SAMPLE AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 97459 | ORTHO SUMMIT SAMPLE HANDLING SYSTEM | SAMPLE PIPETTOR | MMH | HAMILTON BONADUZ AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |