PROMUS ELEMENT ¿
Report
- Report Number
- 2134265-2013-01250
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- November 26, 2012
- Report Date
- February 11, 2013
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KR
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION IDENTIFIED STENT DAMAGE. THE STRUTS BETWEEN THE FIRST ROW ON THE DISTAL END OF THE STENT WERE MISALIGNED. THIS TYPE OF DAMAGE IS CONSISTENT WITH THE DIFFICULTY ENCOUNTERED REMOVING THE DEVICE FROM THE GUIDE CATHETER. THE TIP WAS DAMAGED (THE EDGES WERE JAGGED LIKE). THIS TYPE OF DAMAGE IS CONSISTENT WITH THE TIP BEING PUSHED UP AGAINST A RESTRICTION, DURING ATTEMPTS TO CROSS THE LESION. THE BALLOON WAS VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO ANY POSITIVE PRESSURE. A VISUAL AND MICROSCOPIC EXAMINATION FOUND THAT THE HYPOTUBE WAS KINKED BETWEEN 185MM AND 210MM DISTAL FROM THE CATHETER'S STRAIN RELIEF. NO OTHER KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(6) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION THE DEVICE WAS UNABLE TO CROSS THE LESION. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE LESION WAS PREDILATED WITH A 2.0X15MM MAVERICK BALLOON AND THEN A 3.00X20MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LAD; HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS STABLE. HOWEVER, RETURNED PRODUCT ANALYSIS REVEALED STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 97558 | PROMUS ELEMENT ¿ | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911320300 | 15076306 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |