FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 2992493 · Received March 7, 2013

Report

Report Number
2955842-2013-00757
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
January 11, 2013
Report Date
February 5, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED THE REPORTED BROKEN CABLE. ONE GRIP CABLE WAS BROKEN AT THE DISTAL CLEVIS AND THERE WAS DAMAGE ON IDLER PULLEYS. AROUND THE EDGES OF THE PULLEY, THERE WERE NICKS AND DENTS CREATING A SHARP EDGE WHICH IS LIKELY THE CAUSE OF THE CABLE BREAKING. EVIDENCE NOT CONCLUSIVE, BUT DAMAGE IS LIKELY DUE TO MISHANDLING. THE BROKEN SEGMENTS STUCK OUT APPROXIMATELY .2740 AT THE WRIST. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S PROSTATECTOMY PROCEDURE A CABLE ON THE LARGE NEEDLE DRIVER INSTRUMENT BROKE. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97183 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 420006-06 M10121017 238

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SYSTEM INSTRUMENTS AND ACCESSORIES