FDA Adverse Event Injury Summary report: N

EON MINI

MDR report key: 2992487 · Received March 4, 2013

Report

Report Number
1627487-2013-03314
Event Type
Injury
Date Received
March 4, 2013
Date of Event
September 6, 2012
Report Date
February 6, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-12192011-003-R 1
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

RECALL NUMBER: 1627487-07262012-001-C. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY. THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REFERENCE MFR REPORTS: 1627487-2013-03315. THE PT REPORTED SHE EXPERIENCED HEATING AND BEING BURNED AT HER SCS IPG POCKET SITE WHILE CHARGING HER SCS SYS. THE PT ALSO REPORTED SHE WENT TO THE ER AS A RESULT OF THE BURN. SUBSEQUENTLY, THE PT REDUCED HER CHARGING FREQUENCY. ALSO, A REPLACEMENT CHARGING SYS (LOW ENERGY) WAS SENT TO THE PT. F/U IDENTIFIED THE ISSUE HAS NOT BEEN RESOLVED; IN ADDITION, THE PT IS NO LONGER RECEIVING EFFECTIVE STIMULATION AFTER ENCOUNTERING A METAL DETECTOR. FURTHERMORE, THE PT EXPERIENCED A SHOCK DURING THE ENCOUNTER AND REPORTS HER SCS SYS MALFUNCTIONING FROM THAT POINT. ON (B)(4) 2012 ST. JUDE MEDICAL, NEUROMODULATION DIV, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
91584 EON MINI SCS IPG LGW ST. JUDE MEDICAL - NEUROMODULATION 3788 3108343

Patients

Seq Age Sex Outcome Treatment
1 37 YR Required Intervention IMPLANT:| SCS LEAD: MODEL 3186 (2)