FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2992292 · Received March 7, 2013

Report

Report Number
1416980-2013-05467
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
February 16, 2013
Report Date
February 16, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED FOR EVALUATION; THEREFORE, THE REPORT OF A SYSTEM ERROR 2240 COULD NOT BE CONFIRMED AND THE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. A BATCH REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN TUBING) WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 3 OF 5. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE PATIENT LINE HAD BEEN PROPERLY PRIMED. PATIENT EXTENSION LINES WERE IN USE, AND HAD BEEN PROPERLY CONNECTED PRIOR TO PRIME. THE PATIENT HAD NOT DISCONNECTED PRIOR TO THE ALARM. ALL OF THE BAGS WERE PROPERLY CONNECTED AND THERE WERE NO OPEN CLAMPS ON THE UNUSED LINES. THERE WAS NOTHING UNUSUAL NOTED ABOUT THE SUPPLIES AND THEY WERE NOT DAMAGED BY AN OUTLET PORT CLAMP OR ASSIST DEVICE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HOME PATIENT (HP) CLOSE THE CLAMPS, DISCONNECT, AND CYCLE THE POWER. HC ALARMED SYSTEM ERROR 2367. TSR HAD HP CYCLE POWER AGAIN. HC WENT TO "PRESS GO TO START", AND TSR ASSISTED TO GET SET OUT. PROPER PROCEDURES WERE REVIEWED. PATIENT WOULD START OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT TIME OF INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
97810 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 58 YR HOMECHOICE