FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2992276 · Received March 7, 2013

Report

Report Number
1416980-2013-05465
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
February 15, 2013
Report Date
February 15, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED FOR EVALUATION; THEREFORE, THE REPORT OF A SYSTEM ERROR 2240 COULD NOT BE CONFIRMED AND THE ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. A BATCH REVIEW COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN TUBING) WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL. THE HOME PATIENT (HP) HAD RECEIVED THE ALARM EARLIER AND HAD ALREADY SPOKEN TO HER REGISTERED NURSE. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE PATIENT LINE HAD BEEN PROPERLY PRIMED AND NO PATIENT EXTENSION LINE WERE IN USE. THE PATIENT HAD NOT DISCONNECTED PRIOR. THE HP HAD THREE BAGS CONNECTED AND NO UNUSED LINES. ALL OF THE BAGS WERE PROPERLY CONNECTED. THE HP HAD SOLUTION IN THE HEATER BAG ONLY AND DID NOT NOTICE ANY LEAKS. THERE WAS NOTHING UNUSUAL NOTED ABOUT THE SUPPLIES, AND THEY HAD NOT BEEN DAMAGED BY AN OUTLET PORT CLAMP OR ASSIST DEVICE. THE BAXTER TECHNICAL SERVICES REPRESENTATIVE EXPLAINED THE ALARM. THE PATIENT WOULD COMPLETE THERAPY WITH MANUAL SUPPLIES. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96933 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 51 YR HOMECHOICE