ASAHI APS DIALYZERS
Report
- Report Number
- 8010002-2013-00200
- Event Type
- Injury
- Date Received
- March 1, 2013
- Date of Event
- May 20, 2010
- Report Date
- February 25, 2013
- Manufacturer
- ASAHI KASEI MEDICAL CO., LTD.
- Product Code
- KDI
- PMA / PMN Number
- K051187
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THIS INCIDENT OCCURRED IN (B)(6) AND IS REPORTED TO FDA ACCORDING TO THE REQUIREMENT. APS-15SA IS IDENTICAL MODEL TO REXEED-15S MARKETED IN US. THE USED DEVICE COULD NOT BE ANALYZED BECAUSE IT WAS DISCARDED BY THE USER FACILITY. SO WE REVIEWED MANUFACTURING RECORDS, QUALITY RECORDS OF LOT #301X1A. AS A RESULT, NO ABNORMALITY WAS FOUND IN RECORDS. THE 11,568 UNITS OF THIS LOT # WERE DISTRIBUTED TO THE MEDICAL FACILITIES AND NO SIMILAR EVENT USING THIS LOT # WAS REPORTED GLOBALLY. THE SYMPTOMS OBSERVED IN THE EVENTS ARE CONSIDERED TO BE A DIALYZER REACTION. NOTE: THE DIALYZER REACTION IS A BROAD GROUP OF EVENTS THAT INCLUDE BOTH ANAPHYLACTIC AND LESS WELL- DEFINED ADVERSE REACTIONS OF UNKNOWN CAUSE. "HANDBOOK OF DIALYSIS 4TH ED." JOHN T DAUGIRDAS ET. AL., LIPPINCOTT, WILLIAMS & WILKINS, 2006.
ON (B)(6) 2010: THE DIALYZER (APS-15SA) WAS USED FOR HEMODIALYSIS (HD). AT 15 MINUTES AFTER START OF TREATMENT, BLOOD PRESSURE DECREASED (SYSTOLIC BLOOD PRESSURE: 140=>60MMHG) AND COMPLAINED DYSPNEA. AFTER TACHYCARDIA OCCURRED, DIALYSIS STOPPED TEMPORARILY. AFTER THE ONSET OF THESE ADVERSE EVENTS (AES), PHYSIOLOGICAL SALINE 200ML AND SOLU-CORTEF WERE ADMINISTERED, THEN BLOOD PRESSURE RECOVERED TO 130MMHG. THE DIALYZER WAS CHANGED TO ANOTHER TYPE, THEN HD WAS CONDUCTED WITHOUT PROBLEM. ON (B)(6) 2010: THE DIALYZER (APS-15SA) WAS USED FOR HD. BLOOD PRESSURE DECREASED (SYSTOLIC BLOOD PRESSURE: =>60MMHG) AND PT COMPLAINED CHEST PAIN AND OBSERVED YAWNING EXCESSIVE. AFTER THE ONSET OF THESE AES, PHYSIOLOGICAL SALINE 450ML WAS ADMINISTERED, THEN HD WAS FINISHED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 89467 | ASAHI APS DIALYZERS | KDI | ASAHI KASEI MEDICAL CO., LTD. | APS-15SA | 301X1A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 90 YR | Life Threatening |