FLOGARD
Report
- Report Number
- 1416980-2013-05461
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- February 1, 2013
- Report Date
- February 13, 2013
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4).
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF FAILURE CODE 63 WAS CONFIRMED DURING DEVICE EVALUATION. THE CAUSE WAS IDENTIFIED AS A DAMAGED CENTRAL PROCESSING UNIT (CPU) BOARD. THE CPU BOARD WAS REPLACED TO RESOLVE THE ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A 63 FAILURE CODE. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 97035 | FLOGARD | PUMP, INFUSION | FRN | SHARP CORPORATION OSAKA JAPAN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |