ACCU-CHEK SPIRIT COMBO
Report
- Report Number
- 2183996-2013-00345
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- March 4, 2013
- Report Date
- May 10, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
THE DELIVERY ACCURACY OF THE INSULIN PUMP WAS TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE TECHNICAL INVESTIGATION GAVE NO EVIDENCE THAT THE DEVICE DID CAUSE OR CONTRIBUTE TO THE CONDITION REPORTED BY THE PATIENT.
ON (B)(6) 2013, IT WAS REPORTED THAT THE PATIENT'S INFUSION DEVICE DID NOT BEEP TO CONFIRM HER BOLUS SETTING. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REPLACED AND WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96911 | ACCU-CHEK SPIRIT COMBO | INSULIN INFUSION DEVICE | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 068 YR |