FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT COMBO

MDR report key: 2992015 · Received March 7, 2013

Report

Report Number
2183996-2013-00329
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
August 25, 2012
Report Date
May 23, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATION REGARDING THE CUSTOMER'S ALLEGATION. THE DELIVERY ACCURACY OF THE INSULIN PUMP WAS TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE TECHNICAL INVESTIGATION GAVE NO EVIDENCE THAT THE DEVICE DID CAUSE OR CONTRIBUTE TO THE CONDITION REPORTED BY THE PATIENT. THE PISTON ROD OF THE PUMP WAS TESTED ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE ADAPTER PASSED THE OPTICAL INSPECTION.

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

PATIENT REPORTED EACH TIME SHE CHANGED THE CARTRIDGE FOR THE PAST 6 MONTHS, HER BLOOD GLUCOSE ELEVATED UP TO 600 MG/DL AND REMAINED ELEVATED FOR 24 HOURS. THERE WERE NO AIR BUBBLES IN THE INFUSION SET. SHE BELIEVES THE INSULIN DELIVERY OF THE INFUSION DEVICE IS INACCURATE AND THE PISTON ROD IS "TOO SLOW." SHE SWITCHED TO A DIFFERENT INFUSION DEVICE AND HAS NOT EXPERIENCED FURTHER CONCERNS. SHE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REQUESTED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96872 ACCU-CHEK ® SPIRIT COMBO INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 054 YR