FDA Adverse Event
Malfunction
Summary report: N
PROGRAMMING SOFTWARE
MDR report key: 2991922
·
Received March 6, 2013
Report
- Report Number
- 1644487-2013-00614
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Date of Event
- September 1, 2012
- Report Date
- February 11, 2013
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
REPORTER INDICATED THAT A VNS DELL X5 COMPUTER WITH 8.1 VERSION SOFTWARE WAS FREEZING ON THE 'INTERROGATION SUCCESSFUL' SCREEN. RESEATING THE FLASHCARD DID NOT RESOLVE THE ISSUE. THE COMPUTER AND FLASHCARD HAVE BEEN RETURNED AND ARE PENDING ANALYSIS.
Description of Event or Problem · 1
ANALYSIS OF THE COMPUTER AND FLASHCARD WAS COMPLETED. NO ANOMALIES ASSOCIATED WITH THE HANDHELD COMPUTER WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE COMPUTER PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO SPECIFICATIONS. DURING THE ANALYSIS, NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 94676 | PROGRAMMING SOFTWARE | PROGRAMMING SOFTWARE | LYJ | CYBERONICS, INC. | MODEL 250 | 1073772 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |