FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2991723 · Received March 6, 2013

Report

Report Number
2955842-2013-00754
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
January 24, 2013
Report Date
February 4, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED.  PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE INSTRUMENTS TIPS DID NOT ALIGN. ONE GRIP WAS BENT, CAUSING SIDE TO SIDE MISALIGNMENT OF GRIPS. THERE WAS A .04 OFFSET AT THE TIPS. THE BENT GRIP HAD SEPARATION OF THE YAW PULLEY AND PULLEY COVER AT THE GLUE JOINT, INDICATING OVERLOADING AT THE TIP. EVIDENCE NOT CONCLUSIVE, BUT DAMAGE LIKELY DUE TO MISHANDLING. AN ELECTRICAL CONTINUITY TEST PASSED. ADDITIONAL OBSERVATION FOUND A PITCH CABLE FRAYED AT THE DISTAL CELVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. THE FRAYED SEGMENT IS APPROXIMATELY 0.02 IN LENGTH. NO OTHER CABLE DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING THAT THE TIPS DID NOT ALIGN ON THE MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94924 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT ACCESSORY NAY INTUITIVE SURGICAL,INC. 420172-07 M10121011 236

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM ESU INSTRUMENTS & ACCESSORIES