FDA Adverse Event Malfunction Summary report: N

PROGRASP (TM) FORCEPS INSTRUMENT

MDR report key: 2991696 · Received March 6, 2013

Report

Report Number
2955842-2013-00752
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
January 24, 2013
Report Date
February 6, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS FOUND WITH FRAYED PITCH CABLE LOCATED AT THE DISTAL IDLER WITH LOOSE TENSION. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS INDENTATION DAMAGE ON THE PULLEY. THE INSTRUMENT EXHIBITED INDENTATIONS AND SCRATCH MARKS ON THE DISTAL PULLEY SURFACE. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS MAIN TUBE DAMAGE. THE DISTAL END OF THE MAIN TUBE EXHIBITED VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND WERE NOT ALIGNED WITH THE TUBE AXIS. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI PROSTATECTOMY PROCEDURE, THE CUSTOMER NOTED THAT THE WIRES WERE OFF TRACK ON THE PROGRASP FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.