PROGRASP (TM) FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-00752
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Date of Event
- January 24, 2013
- Report Date
- February 6, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE INSTRUMENT WAS FOUND WITH FRAYED PITCH CABLE LOCATED AT THE DISTAL IDLER WITH LOOSE TENSION. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS INDENTATION DAMAGE ON THE PULLEY. THE INSTRUMENT EXHIBITED INDENTATIONS AND SCRATCH MARKS ON THE DISTAL PULLEY SURFACE. ADDITIONAL OBSERVATION NOT REPORTED BY SITE WAS MAIN TUBE DAMAGE. THE DISTAL END OF THE MAIN TUBE EXHIBITED VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND WERE NOT ALIGNED WITH THE TUBE AXIS. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS.
IT WAS REPORTED THAT DURING A DAVINCI SI PROSTATECTOMY PROCEDURE, THE CUSTOMER NOTED THAT THE WIRES WERE OFF TRACK ON THE PROGRASP FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 95911 | PROGRASP (TM) FORCEPS INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420093-08 | M10120831 155 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |