FDA Adverse Event Malfunction Summary report: N

VITROS 5600 INTEGRATED SYSTEM

MDR report key: 2991666 · Received March 6, 2013

Report

Report Number
1319681-2013-00059
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
February 5, 2013
Report Date
March 6, 2013
Manufacturer
ORTHO-CLINICAL DIAGNOSTICS
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION DETERMINED THAT MULTIPLE, NON-REPRODUCIBLE, HIGHER THAN EXPECTED, VITROS TROP I ES RESULTS WERE OBTAINED FROM TWO DIFFERENT SAMPLES OBTAINED FROM A SINGLE PATIENT PROCESSED ON THE VITROS 5600 SYSTEM. SERVICE REPAIR ACTIONS WERE PERFORMED TO RETURN THE SYSTEM TO NOMINAL OPERATION. ADDITIONALLY, THE SAMPLE MAY NOT HAVE BEEN PROCESSED IN ACCORDANCE WITH THE TUBE MANUFACTURER'S RECOMMENDATIONS. IT IS LIKELY THAT CELLULAR DEBRIS, DUE TO POOR SAMPLE PREPARATION, WAS PRESENT IN THE AFFECTED SAMPLE, ALTHOUGH THIS COULD NOT BE CONFIRMED. THE MOST LIKELY ASSIGNABLE CAUSE IS AN INSTRUMENT RELATED ISSUE. ADDITIONALLY, A PRE-ANALYTICAL SAMPLE PROCESSING ISSUE CANNOT BE RULED OUT AS A CONTRIBUTING FACTOR. THERE WAS NO EVIDENCE THAT A REAGENT RELATED ISSUE CONTRIBUTED TO THE EVENT.

Description of Event or Problem · 1

THE CUSTOMER OBTAINED MULTIPLE, NON-REPRODUCIBLE, HIGHER THAN EXPECTED, VITROS TROP I ES RESULTS (SAMPLE 1 = 0.070, SAMPLE 2 = 0.063, 0.064 VS. EXPECTED RESULT = 0.028 NG/ML) FROM TWO DIFFERENT SAMPLES OBTAINED FROM A SINGLE PATIENT PROCESSED ON THE VITROS 5600 SYSTEM. THE AFFECTED PATIENT RESULT OF 0.063 NG/ML WAS REPORTED OUT OF THE LABORATORY. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THE CUSTOMER REPEATED THE AFFECTED SAMPLES AS THE PHYSICIAN QUESTIONED THE RESULT. A CORRECTED RESULT WAS ISSUED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95947 VITROS 5600 INTEGRATED SYSTEM CHEMISTRY ANALYZER MMI ORTHO-CLINICAL DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1