FDA Adverse Event Malfunction Summary report: N

GOLD RELOAD FOR ECHELON

MDR report key: 2991583 · Received March 6, 2013

Report

Report Number
3005075853-2013-01108
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
February 11, 2013
Report Date
February 20, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PREMATURE SLED MOVEMENT. THE ANALYSIS RESULTS SHOWED THAT ONE RELOAD WAS RECEIVED PARTIALLY FIRED 1/16. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. THE RETURNED RELOAD WAS RESET AND LOADED INTO A TEST DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION AND ACHIEVED ITS COMPLETE STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. NO MALFORMED STAPLES WERE NOTED DURING FIRING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SLEEVE GASTRECTOMY PROCEDURE, THE RELOAD DID NOT FORM THE STAPLE. ANOTHER LIKE RELOAD WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94793 GOLD RELOAD FOR ECHELON STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA J4CN12

Patients

Seq Age Sex Outcome Treatment