GOLD RELOAD FOR ECHELON
Report
- Report Number
- 3005075853-2013-01108
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Date of Event
- February 11, 2013
- Report Date
- February 20, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051002
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER
Narratives
(B)(4). PREMATURE SLED MOVEMENT. THE ANALYSIS RESULTS SHOWED THAT ONE RELOAD WAS RECEIVED PARTIALLY FIRED 1/16. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. THE RETURNED RELOAD WAS RESET AND LOADED INTO A TEST DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION AND ACHIEVED ITS COMPLETE STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. NO MALFORMED STAPLES WERE NOTED DURING FIRING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
IT WAS REPORTED THAT DURING A SLEEVE GASTRECTOMY PROCEDURE, THE RELOAD DID NOT FORM THE STAPLE. ANOTHER LIKE RELOAD WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 94793 | GOLD RELOAD FOR ECHELON | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4CN12 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |