FDA Adverse Event Injury Summary report: N

FLEXICAP

MDR report key: 2991567 · Received March 6, 2013

Report

Report Number
1416980-2013-05388
Event Type
Injury
Date Received
March 6, 2013
Date of Event
February 9, 2013
Report Date
February 13, 2013
Manufacturer
BAXTER HEALTHCARE - SWINFORD
Product Code
KDJ
PMA / PMN Number
K972579
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT OF PERITONITIS WAS RECEIVED WITH NO ALLEGED DEVICE MALFUNCTION OR USE ERROR; THEREFORE, A SAMPLE WAS NOT REQUESTED. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR ASSOCIATED LOT NUMBER 12J16H25 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. THE PROBLEM WAS NOT CONFIRMED. NO DEVICE MALFUNCTION OR USE ERROR WAS IDENTIFIED. NO ASSIGNABLE CAUSE WAS DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT EXPERIENCED PERITONITIS COINCIDENT WITH THERAPY FOR PERITONEAL DIALYSIS (PD). THE PATIENT EXPERIENCED A COUGH AND WAS DIAGNOSED WITH BRONCHITIS. THE CAUSE OF THE BRONCHITIS WAS THE PATIENT'S MEDICAL HISTORY OF CHRONIC OBSTRUCTIVE PULMONARY DISEASE (COPD). THE PATIENT WAS TREATED FOR THE COUGH AND BRONCHITIS WITH PREDNISONE AND UNSPECIFIED ANTIBIOTICS. ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED TRAUMA TO THEIR PD CATHETER EXIT SITE, CAUSED BY THEIR COUGH, AND THE PD CATHETER MIGRATED. ON A LATER DATE, THE PATIENT EXPERIENCED PERITONITIS AND WAS ADMITTED TO THE HOSPITAL. THE CAUSE OF THE PERITONITIS WAS UNKNOWN. TREATMENT INFORMATION WAS NOT REPORTED. THE PATIENT'S PD CATHETER WAS PULLED AND PD THERAPIES WERE DISCONTINUED. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL AND HAS RECOVERED FROM THE EVENTS OF COUGH, BRONCHITIS, TRAUMA TO THE PD CATHETER EXIT SITE AND PERITONITIS. THIS IS THE SAME PATIENT AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95472 FLEXICAP SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - SWINFORD

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization HOMECHOICE, DIANEAL AMBUFLEX, DIANEAL ULTRABAG| EXTRANEAL