ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-02435
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Report Date
- February 20, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION 08/28/2015-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/10/2015 WITH THE FOLLOWING FINDINGS: A VISUAL INSPECTION OF THE PUMP SHOWED THAT THE KEYPAD AUDIO BOLUS BUTTON WAS UNDAMAGED. THE AUDIO BOLUS BUTTON RESPONDED APPROPRIATELY DURING TESTING. THE COMPLAINT WAS NOT DUPLICATED. UNRELATED TO THE COMPLAINT, THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE KEYPAD COVER WAS REMOVED AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL THE KEY CONTACTS. ADDITIONALLY, THE DISPLAY SCREEN WAS DIM. THE BATTERY COMPARTMENT WAS CRACKED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. IT WAS REPORTED THAT THE AUDIO BOLUS BUTTON WAS INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE HARD PRESSES TO ELICIT A RESPONSE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 95084 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |