FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 2991342 · Received March 6, 2013

Report

Report Number
2955842-2013-00742
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
December 4, 2013
Report Date
February 5, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT CABLE WAS BROKEN. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINED THE CRIMP WAS STILL INSTALLED IN CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. THE ELECTRICAL CONTINUITY TESTING PASSED. ENGINEERING ALSO FOUND SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED DURING A DA VINCI SI PROSTATECTOMY PROCEDURE THAT THE FENESTRATED BIPOLAR INSTRUMENT CABLE BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95318 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120104 120

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES