FDA Adverse Event Malfunction Summary report: N

PWRD ECH FLEX 60MM

MDR report key: 2991299 · Received March 6, 2013

Report

Report Number
3005075853-2013-01086
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
February 1, 2013
Report Date
February 8, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC.)? WHERE THERE STAPLES SEEN IN THE SURGICAL FIELD? IF STAPLES WERE SEEN IN THE SURGICAL FIELD, WHAT WAS THERE APPEARANCE? WHAT COLOR CARTRIDGE WAS BEING USED? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE?

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE (B)(4) DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE (B)(4) CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED AND IN GOOD VISUAL CONDITION. THE RECEIVED CARTRIDGE WAS DISASSEMBLED TO INSPECT THE INTERNAL COMPONENTS AND NO ANOMALIES WERE FOUND. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY PROCEDURE, THE SURGEON STATED TO ONE OF HIS HBT THAT HE APPLIED STAPLER TO TISSUE AND FIRED THE DEVICE AND IT APPEARED THAT THE STAPLER CUT WITHOUT STAPLING. THE REP WAS TOLD THE CASE WAS COMPLETED AND THE PATIENT IS DOING FINE. THERE WERE NO PATIENT CONSEQUENCES. UNKNOWN HOW CASE WAS COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95237 PWRD ECH FLEX 60MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1