FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPSINSTRUMENT

MDR report key: 2991296 · Received March 6, 2013

Report

Report Number
2955842-2013-00743
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
January 10, 2013
Report Date
February 5, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT CABLE WAS BROKEN. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINED THE CRIMP STILL REMAINED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PANCREATECTOMY (WHIPPLE PROCEDURE) PROCEDURE THE FENESTRATED BIPOLAR INSTRUMENT CABLE BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95236 FENESTRATED BIPOLAR FORCEPSINSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120530 598

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES