FENESTRATED BIPOLAR FORCEPSINSTRUMENT
Report
- Report Number
- 2955842-2013-00743
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Date of Event
- January 10, 2013
- Report Date
- February 5, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT CABLE WAS BROKEN. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINED THE CRIMP STILL REMAINED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI PANCREATECTOMY (WHIPPLE PROCEDURE) PROCEDURE THE FENESTRATED BIPOLAR INSTRUMENT CABLE BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 95236 | FENESTRATED BIPOLAR FORCEPSINSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10120530 598 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |