FDA Adverse Event Injury Summary report: N

EXPEDIUM 5.50 DUAL-INNIE POLYAXIAL SCREW 8.00X100MM, TI

MDR report key: 2991239 · Received March 6, 2013

Report

Report Number
1526439-2013-13245
Event Type
Injury
Date Received
March 6, 2013
Date of Event
February 5, 2013
Report Date
April 12, 2013
Manufacturer
DEPUY SYNTHES SPINE
Product Code
NKB
PMA / PMN Number
PK033901
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED UPON RECEIPT OF THE DEVICE AND COMPLETION OF THE EVALUATION.

Additional Manufacturer Narrative · 1

A VISUAL INSPECTION CONFIRMED THAT THE EXPEDIUM DUAL INNIE POLYAXIAL BROKE AT THE TRANSITION FROM THE SCREW HEAD TO THE SCREW SHANK. A SCANNING ELECTRON MICROSCOPY (SEM) ANALYSIS WAS PERFORMED ON THE FRACTURED SURFACES TO FACILITATE A MORE DETAILED INVESTIGATION OF THE FRACTURE CHARACTERISTICS OF THE DAMAGED SAMPLE. RESULTS ACKNOWLEDGED AN EVIDENT FATIGUE STRIATION WHERE THE CRACK TERMINATION TOOK PLACE. FURTHERMORE, THE EXISTENCE OF THE TWO SURFACE MORPHOLOGIES ILLUSTRATED THAT THE FRACTURE FIRST PROPAGATED AND THEN A FULL FAILURE/BREAKAGE TOOK PLACE PRESUMABLY WHEN THE REMAINING REGION DID NOT HAVE STRENGTH TO BEAR THE LOAD. NO MATERIAL DEFECTS OR OTHER ABNORMALITIES WERE OBSERVED IN THIS ANALYSIS. THE ROOT CAUSE CANNOT BE POSITIVELY DETERMINED. HOWEVER, THE SEM FRACTURE ANALYSIS SUGGESTS THAT THE SCREW FRACTURE MOST LIKELY OCCURRED DUE TO FATIGUE. IT WAS REPORTED THAT PATIENT IS A SMOKER AND OVERWEIGHT. IN THE ABSENCE OF A PRODUCT ISSUE OR OBSERVED TREND, THIS COMPLAINT WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED.

Description of Event or Problem · 1

REVISION SURGERY WAS PERFORMED TO REMOVE TWO BROKEN SCREWS. PATIENT HAD COMPLAINED OF PAIN AND X-RAYS NOTED BROKEN HARDWARE. SEE MFG MEDWATCH REPORT NO. 1526439-2013-13246 FOR THE SECOND SCREW THAT WAS INVOLVED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95094 EXPEDIUM 5.50 DUAL-INNIE POLYAXIAL SCREW 8.00X100MM, TI ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE NKB DEPUY SYNTHES SPINE RL146546

Patients

Seq Age Sex Outcome Treatment
1 55 YR Required Intervention