FDA Adverse Event Injury Summary report: N

PINNACLE SECTOR II CUP 52MM

MDR report key: 2990960 · Received March 6, 2013

Report

Report Number
1818910-2013-13232
Event Type
Injury
Date Received
March 6, 2013
Date of Event
February 21, 2013
Report Date
January 28, 2015
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
LPH
PMA / PMN Number
PK000306
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODES REQUIRED WERE NOT PROVIDED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. D. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITH THE INFORMATION AVAILABLE. BASED ON THE INABILITY TO IDENTIFY A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. REVIEW OF THE PROVIDED PRODUCT/LOT DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. MEDICAL RECORDS WERE REVIEWED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED.

Additional Manufacturer Narrative · 1

THE COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. UPDATE: LITIGATION PAPERS RECEIVED 1/2/2014. LITIGATION ALLEGES THAT THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, INFLAMMATION, AND LARGE AMOUNTS OF TOXIC COBALT CHROMIUM METAL ION PARTICLES TO BE RELEASED INTO THE BLOOD, TISSUE, AND BONE. DATE OF IMPLANT HAS ALSO BEEN PROVIDED AND PART/LOT INFORMATION LOCATED THROUGH AN INVOICE SEARCH. A FEMORAL HEAD IS NOW BEING REPORTED TO ADDRESS THESE ALLEGATIONS. THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS DID NOT REVEAL ANY RELATED MANUFACTURING DEVIATIONS OR ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

THE PATIENT WAS REVISED BECAUSE OF CUP POSITION, BROKEN SCREW, OSTEOLYSIS, AND POLY WEAR OF THE LINER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
96069 PINNACLE SECTOR II CUP 52MM CUP LPH DEPUY ORTHOPAEDICS INC US YS1ET1000

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention