FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 2990905 · Received March 6, 2013

Report

Report Number
2134265-2013-01334
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
December 20, 2012
Report Date
February 8, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GB
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATED BY MFR: RETURNED PRODUCT CONSISTED OF A NC QUANTUM APEX MONORAIL (MR) BALLOON CATHETER WITH NO ORIGINAL PACKAGING OR OTHER DEVICES. THERE WAS A COMPLETE SEPARATION OF THE HYPOTUBE. THE HYPOTUBE SEPARATION WAS 71.5 CM FROM THE HUB. THE HYPOTUBE FRACTURE SURFACES WERE OVALED, WHICH SUGGESTS THE DEVICE WAS KINKED PRIOR TO SEPARATION. THERE WAS CONTRAST IN THE INFLATION LUMEN, WIRE LUMEN AND ON THE OUTSIDE OF THE CATHETER AND BALLOON. THE BALLOON WAS TIGHTLY FOLDED. THE HYPOTUBE WAS KINKED 3.5 CM FROM THE FRACTURED DISTAL END. THERE IS NO EVIDENCE OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES CONTRIBUTING TO THE REPORTED EVENT. PRODUCT ANALYSIS CONFIRMED THE REPORTED KINK AND CONFIRMED A BROKEN HYPOTUBE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(4) 2013. IT WAS REPORTED THAT DURING PREPARATION FOR AN ANGIOPLASTY TREATMENT PROCEDURE, THE 3.0 X 20 MM NC QUANTUM APEX BALLOON WAS FOUND TO BE KINKED IN 2 PLACES. HOWEVER, THE PRODUCT ANALYSIS REVEALED A SHAFT FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95898 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912420300 15654435

Patients

Seq Age Sex Outcome Treatment
1