ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2013-00327
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Date of Event
- March 4, 2013
- Report Date
- April 16, 2013
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE MOISTURE ENTERED THE INSULIN PUMP AND DESTROYS THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON IS PERMANENT ACTIVE DUE TO EXTERNAL MECHANIC DAMAGE. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
ON (B)(6) 2013, IT WAS REPORTED THAT ALL OF THE BUTTONS ON THE PATIENT'S INFUSION DEVICE WERE NO LONGER FUNCTIONING. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REPLACED AND WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 95029 | ACCU-CHEK SPIRIT | INSULIN INFUSION DEVICE | LZG | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |