FDA Adverse Event Malfunction Summary report: N

D/M CRIMP TOOL

MDR report key: 2990825 · Received March 6, 2013

Report

Report Number
0002249697-2013-00914
Event Type
Malfunction
Date Received
March 6, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

VISUAL EVALUATION SHOWED THE DOWEL PIN WAS MISSING FROM THE RATCHET ASSEMBLY, AND THERE WAS NO SIGN OF A WELD. MATERIAL ANALYSIS CONCLUDED THAT THERE WAS NO EVIDENCE OF INTERFERENCE BETWEEN ON THE RATCHET ASSEMBLY HOLES EXAMINED. DIMENSIONAL ANALYSIS DETERMINED THAT THE LEVER PIN HOLE WAS OVERSIZED. THE EVENT WAS CONFIRMED. A DEVICE HISTORY REVIEW INDICATED THAT ALL DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW INDICATED THAT THERE HAVE BEEN NO SIMILAR EVENTS FOR THE REPORTED LOT. THE INVESTIGATION CONCLUDED THAT THE DOWEL PIN COMING OUT OF THE RATCHET ASSEMBLY OF THE D/M CRIMP TOOL WAS CAUSED BY A MANUFACTURING NONCONFORMANCE. BASED UPON THE CONCLUSIONS OF THE VISUAL, MATERIAL, AND DIMENSIONAL ANALYSIS, IT WAS CONCLUDED THAT THE SUPPLIER, (B)(4), HAD NOT PERFORMED THE REQUIRED WELDING OPERATION ON THE RATCHET ASSEMBLY AND HAD MANUFACTURED THE LEVER PIN HOLE OVERSIZED WHICH BOTH LED TO THE PIN COMING OUT OF THE ASSEMBLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RATCHET OF THE CRIMP TOOL WAS NOT FUNCTIONAL IN THA. THE SURGEON NOTICED THE PIN OF THE RATCHET WAS OFF AND PIN WAS NOT FOUND. SO, THE SURGEON CHECKED THE X-RAY AND THERE WAS NO PIN IN THE X-RAY. WE COULD NOT SPECIFY WHEN THE PIN WAS MISSED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RATCHET OF THE CRIMP TOOL WAS NOT FUNCTIONAL IN THA. THE SURGEON NOTICED THE PIN OF THE RATCHET WAS OFF AND PIN WAS NOT FOUND. SO, THE SURGEON CHECKED THE X-RAY AND THERE WAS NO PIN IN THE X-RAY. WE COULD NOT SPECIFY WHEN THE PIN WAS MISSED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
95799 D/M CRIMP TOOL INSTRUMENT LXH STRYKER ORTHOPAEDICS-MAHWAH RD6S186

Patients

Seq Age Sex Outcome Treatment
1 Other