FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2990818 · Received March 5, 2013

Report

Report Number
2531779-2013-02395
Event Type
Malfunction
Date Received
March 5, 2013
Report Date
February 12, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING EVALUATION, THE BATTERY COMPARTMENT THREADS WERE STRIPPED AND THE CAP WAS GETTING STUCK. THE BATTERY CAP THREADS WERE STRIPPED AND THE BATTERY CAP HEIGHT AND WIDTH WERE OUT OF SPECIFICATIONS. THE O-RING WAS MISSING. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE BATTERY CAP HEIGHT AND WIDTH WERE OUT OF SPECIFICATIONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93768 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 42 YR