FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2990818
·
Received March 5, 2013
Report
- Report Number
- 2531779-2013-02395
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Report Date
- February 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING EVALUATION, THE BATTERY COMPARTMENT THREADS WERE STRIPPED AND THE CAP WAS GETTING STUCK. THE BATTERY CAP THREADS WERE STRIPPED AND THE BATTERY CAP HEIGHT AND WIDTH WERE OUT OF SPECIFICATIONS. THE O-RING WAS MISSING. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACKED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE BATTERY CAP HEIGHT AND WIDTH WERE OUT OF SPECIFICATIONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93768 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |