COULTER® AC*T DIFF ANALYZER
Report
- Report Number
- 1061932-2013-00355
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Date of Event
- February 13, 2013
- Report Date
- February 13, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K973634
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
LV7 IS A THREE-WAY SOLENOID VALVE USED TO ROUTE DILUENT DISPENSED BY THE SAMPLE SYRINGE VIA LV11 TO THE LOWER-SIDE PORTS OF THE BATHS. LV11 IS A THREE-WAY SOLENOID VALVE USED TO CONTROL DILUENT OUTPUT FROM THE SAMPLE SYRINGE. LV12 IS A THREE-WAY SOLENOID VALVE USED TO CONTROL THE ROUTING OF DILUENT. (B)(4).
THE CUSTOMER REPORTED A LEAK OF DILUENT INVOLVING THE COULTER ACT DIFF ANALYZER. THE CUSTOMER INDICATED THAT THE LEAK WAS APPROXIMATELY 10 ML IN VOLUME AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE EVENT AND NO INJURY OR EXPOSURE WAS REPORTED. NO ERRONEOUS RESULTS WERE GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2013 AND NOTED THAT THERE WAS A FLUID LEAK BUT WAS UNABLE TO FIND THE SOURCE OF THE LEAK. THE FSE REPLACED THE PINCH VALVES FOR LV12, LV11, AND LV7 AS A PREVENTIVE MEASURE. THE FSE THEN PERFORMED INSTRUMENT STARTUP AND RAN QUALITY CONTROL (QC) WITHOUT ANY FURTHER LEAKS. THE CUSTOMER HAS NOT REPORTED A RECURRENCE OF THIS EVENT AS OF (B)(6) 2013. FAILURE MODE WAS DETERMINED TO BE ONE, OR A COMBINATION OF THE PINCH VALVES REPLACED DURING SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 92850 | COULTER® AC*T DIFF ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |