FDA Adverse Event Malfunction Summary report: N

REMB SAG SAW

MDR report key: 2990779 · Received March 5, 2013

Report

Report Number
0001811755-2013-00451
Event Type
Malfunction
Date Received
March 5, 2013
Date of Event
February 18, 2013
Report Date
February 18, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE MOTOR ASSEMBLY INCLUDING THE PRINTED CIRCUIT BOARD (TPS FLEX ASSEMBLY) WERE FOUND TO BE CORRODED. THE PRESENCE OF CORROSION WAS IDENTIFIED AS A PROBABLE CAUSE OF THE REPORTED BIAS CURRENT ERROR, AS CORROSION CAN INTERFERE WITH THE DEVICE'S ELECTRICAL CURRENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE SAW CAUSED A BIAS CURRENT WARNING TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93631 REMB SAG SAW DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1