FDA Adverse Event
Malfunction
Summary report: N
RESERVOIR 1.8ML
MDR report key: 2990690
·
Received March 5, 2013
Report
- Report Number
- 3004209178-2013-91627
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Date of Event
- February 23, 2013
- Report Date
- February 23, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K001828
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Additional Manufacturer Narrative · 1
INSPECTED FOUR OPENED AND USED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED PAST THE O-RINGS AND INTO THE RESERVOIR COMPARTMENT. THE BLOOD GLUCOSE READING WAS 156MG/DL. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 94177 | RESERVOIR 1.8ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-326A | H8486688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |