FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 2990530 · Received March 5, 2013

Report

Report Number
2134265-2013-01528
Event Type
Malfunction
Date Received
March 5, 2013
Date of Event
February 19, 2013
Report Date
February 20, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED CALCIFIED LESION WAS LOCATED IN THE MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING (LAD) ARTERY. AFTER CROSSING THE LESION WITH A 5.00 X 8 MM NC QUANTUM APEX BALLOON, THE DEVICE WAS INFLATED. DURING INFLATION THE BALLOON RUPTURED AT 20 ATMS. IT IS UNKNOWN HOW MANY INFLATIONS WERE PERFORMED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93621 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912408500 15673351

Patients

Seq Age Sex Outcome Treatment
1