FDA Adverse Event Malfunction Summary report: N

ADULT UNHEATED BREATHING CIRCUIT

MDR report key: 2990462 · Received March 5, 2013

Report

Report Number
9611451-2013-00159
Event Type
Malfunction
Date Received
March 5, 2013
Date of Event
January 31, 2013
Report Date
February 7, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
K033710
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE DRYLINE TUBES OF THE COMPLAINT RT134 ADULT BREATHING CIRCUITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT BOTH DRYLINE TUBES SUSTAINED A HOLE, ABOUT 17.5CM AWAY FROM THE CONNECTOR. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 120305. CONCLUSION: IT WAS IDENTIFIED THAT DAMAGE TO THE RT134 DRYLINES WAS CAUSED BY A FAULTY CORRUGATOR BLOCK THAT WAS USED DURING THE DRYLINE MANUFACTURING PROCESS. THE FAULTY BLOCK WAS LAST YEAR. THE COMPLAINT DRYLINE TUBES WERE MANUFACTURED PRIOR TO THE REPLACEMENT OF THE FAULTY CORRUGATOR BLOCK. DURING THE PRODUCTION OF THE DRYLINE TUBES, A PRESSURE TEST IS PERFORMED EVERY (B)(4) AND THOSE THAT FAIL ARE SET ASIDE FOR FURTHER INSPECTION. THE USER INSTRUCTIONS FOR THE RT134 ADULT UNHEATED BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA A DISTRIBUTOR THAT TWO RT134 ADULT UNHEATED BREATHING CIRCUITS FAILED THE VENTILATOR LEAK TEST. IT WAS FURTHER REPORTED THAT A PINHOLE WAS FOUND ON EACH OF THE DRYLINE TUBE. THIS WAS OBSERVED BEFORE PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
93616 ADULT UNHEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT134 120305

Patients

Seq Age Sex Outcome Treatment
1