DIMENSION® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2013-00102
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Date of Event
- March 28, 2012
- Report Date
- February 8, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- JFY
- PMA / PMN Number
- K073055
- Removal / Correction Number
- 2517506-02-22-2013-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
SIEMENS HEALTHCARE DIAGNOSTICS INC. HAS CONFIRMED CUSTOMER COMPLAINTS OF FALSELY DEPRESSED ENZYMATIC CREATININE (EZCR) RESULTS WHEN EZCR IS PROCESSED FROM OPEN WELLS OF ENZYMATIC CREATININE (EZCR) FLEX REAGENT CARTRIDGES THAT ARE IN CLOSE PROXIMITY TO OPEN WELLS OF PHOSPHORUS (PHOS) FLEX REAGENT CARTRIDGES. THE FALSELY DEPRESSED EZCR RESULTS ARE CAUSED BY A EZCR REAGENT INTERACTION WITH THE PHOS REAGENT. AN URGENT MEDICAL DEVICE CORRECTION FOR THE PHOS FLEX REAGENT CARTRIDGE (DF61), COMMUNICATION #13-03 WAS ISSUED IN FEBRUARY 2013 TO IMPACTED CUSTOMERS. THE COMMUNICATION PROVIDED REMEDIAL ACTIONS TO CUSTOMERS TO EITHER AVOID THE REAGENT INTERACTIONS OR TO DISCONTINUE THE USE OF EITHER PHOS OR EZCR ON THEIR DIMENSION SYSTEM.
FALSELY DEPRESSED ENZYMATIC CREATININE RESULTS WERE OBTAINED ON QC SAMPLES. IT IS UNKNOWN IF PATIENT RESULT WERE REPORTED TO THE PHYSICIAN. THE SAMPLES WERE REPEATED WITH A NEW REAGENT SEGMENT (WELL) AND HIGHER RESULTS WERE OBTAINED. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED AS A RESULT OF THE FALSELY DEPRESSED CREATININE QC RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED CREATININE QC RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93846 | DIMENSION® CLINICAL CHEMISTRY SYSTEM | ENZYMATIC CREATININE FLEX® REAGENT CARTRIDGE | JFY | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | GA2351 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |