PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2013-01281
- Event Type
- Injury
- Date Received
- March 5, 2013
- Date of Event
- February 6, 2013
- Report Date
- February 12, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE PATIENT WAS REPORTED TO BE MORBIDLY OBESE WITH A BODY MASS INDEX OF (B)(6). PER THE SPECIAL PATIENT POPULATIONS SECTION OF THE PERCLOSE PROGLIDE DEVICE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PERCLOSE PROGLIDE DEVICE HAS NOT BEEN ESTABLISHED IN THE PATIENT POPULATIONS WHO ARE MORBIDLY OBESE WITH A BODY MASS INDEX GREATER THAN (B)(6). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED NEEDLE-TO-CUFF MISS WAS CONFIRMED. BASED ON VISUAL INSPECTION AND FUNCTIONAL TESTING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT PRIOR TO AN ABDOMINAL AORTIC ANEURYSM (AAA) REPAIR PROCEDURE, PRE-CLOSE PLACEMENT OF THE SUTURES WAS ACHIEVED OF THE RIGHT COMMON FEMORAL ARTERY USING PERCLOSE PROGLIDE DEVICES THROUGH A 6-FRENCH SIZED ACCESS SITE. REPORTEDLY, DURING DEPLOYMENT OF THE INITIAL PROGLIDE DEVICE A NEEDLE-TO-CUFF MISS OCCURRED. THE DEVICE WAS REMOVED OVER A GUIDE WIRE AND THE SUTURES FROM TWO ADDITIONAL PROGLIDE DEVICES WERE SEQUENTIALLY DEPLOYED 60 DEGREES OPPOSITE IN ORIENTATION AND SET TO THE SIDE. THE ACCESS SITE WAS DILATED TO 18-FRENCH TO MATCH THE OUTER DIAMETER OF THE DELIVERY CATHETER. AFTER CONCLUSION OF THE AAA REPAIR PROCEDURE, THE KNOTS FROM THE TWO PROGLIDE DEVICES WERE SEQUENTIALLY ADVANCED AND TIGHTENED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO REPORTED SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THE PHYSICIAN WAS REPORTED TO BE TRAINED IN THE USE OF THE PERCLOSE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 94010 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20604J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention | SHEATH: 6-FRENCH |