ULTRA FF
Report
- Report Number
- 1219602-2013-00058
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Date of Event
- January 8, 2013
- Report Date
- February 6, 2013
- Manufacturer
- MANSFIELD MANUFACTURING SITE
- Product Code
- GAT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
VISUAL INSPECTION CONFIRMED THAT T1 WAS NOT RETURNED, AS IT WAS REPORTED THAT IT WAS LEFT IN THE PATIENT. THE SUTURE WAS OBSERVED TO BE CUT RIGHT AT THE BOTTOM OF THE T1 LOCATION, THIS OBSERVATION INDICATES THAT THE SUTURE MAY HAVE BEEN CUT BY THE NEEDLE AS THE SURGEON STATED HE/SHE TURNED THE NEEDLE 90 DEGREES AND PULLED IT OUT UPON IMPLANTING T1. FUNCTIONAL TESTING WAS PERFORMED AND THE ACTUATOR WAS FULLY ADVANCED OVER THE DIMPLE AND T2 DEPLOYED WITHOUT ISSUE INTO A RUBBER MENISCUS, FUNCTIONING AS INTENDED. BASED ON THESE OBSERVATIONS, NO ROOT CAUSE RELATED TO THE MANUFACTURE OF THE DEVICE CAN BE ESTABLISHED. NO FURTHER INVESTIGATION IS WARRANTED AT THIS TIME. (B)(4).
WHEN THE SURGEON INSERTED THE ULTRA FAST-FIX INTO THE MENISCUS AND TURNED IT 90 DEGREES AND PULLED OUT TO DEPLOY THE T1 IMPLANT, THEY FOUND THAT THE SUTURE BROKE AWAY FROM THE T1 AND THE KNOT SEEMED TO BE UNTIED. ACCORDING THE REPORT FROM THE SURGEON, THEY USED THE DEVICE AS USUAL AND DID NOT PULL WITH EXCESSIVE FORCE. THE T1 REMAINS IN THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 93839 | ULTRA FF | ULTRA FAST-FIX ASSEMBLY -CVD | GAT | MANSFIELD MANUFACTURING SITE | 72201491 | 50401406 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |