LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2013-01049
- Event Type
- Malfunction
- Date Received
- March 5, 2013
- Date of Event
- January 29, 2013
- Report Date
- February 20, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE ANALYSIS RESULT SHOWED THAT THE INSTRUMENT WAS RECEIVED WITH THE JAWS FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE RETURNED CONDITION OF THE DEVICE. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS WERE NOT COMPRESSING. THE MALFUNCTION ADDED FIFTEEN MINUTES TO THE CASE. NO ADDITIONAL BLOOD LOSS DUE TO THE MALFUNCTION. IT IS UNKNOWN HOW THE CASE WAS COMPLETED. THIS IS ALL THE INFORMATION AVAILABLE.